mobile-process
 
 

On-line TOC, BORON Analyzers

Sievers 860 Laboratory Total Organic Carbon (TOC) Analyzer

Sievers 860 Laboratory Total Organic Carbon (TOC) Analyzer

REQUEST QUOTATION

The Sievers 860 Laboratory Total Organic Carbon (TOC) Analyzer was designed specifically for pharmaceutical applications. Based on the proven technology used in the best-selling Sievers 900 Laboratory model, the new Sievers 860 provides excellent analytical performance and a streamlined feature set designed to meet general pharmaceutical water testing requirements, including purified water (PW) and water for injection (WFI). It has a linear range of 0.3 ppb to 1.5 ppm, and offers a more robust alternative to the direct conductivity method.

Like other Sievers TOC analyzers, the Sievers 860 is easy to set up, operate and maintain, and requires no special operator training. To keep cost of ownership low, the Sievers 860 needs no internal reagents. The Analyzer offers automated calibration and validation, fast analysis, and automated reagent adjustment. For demanding lab applications, the 860 can be connected to the optional GE Autosampler RT12.

The innovative DataPlus software makes the Sievers 860 more than an analyzer. DataPlus gives operators new insight into water system performance. Featuring an intuitive PC-based interface, the software lets you easily manage the data sample points in your water system and comply with global TOC monitoring requirements, such as the US, European, Chinese, Indian, and Japanese Pharmacopeias (USP, EP, ChP, IP, and JP).

Key features and benefits of the Sievers 860 Laboratory Total Organic Carbon (TOC) Analyzer include:

Proven Sievers technology

The Sievers 860 Laboratory TOC Analyzer uses the proven Sievers Membrane Conductometric TOC technology that provides unsurpassed analytical performance and reliability.

Regulatory Compliance

The Sievers 860 complies with major global pharmacopeia requirements.

Affordability

The Sievers 860's streamlined feature set was carefully designed to reduce costs, while still offeringhigh performance for select pharmaceutical applications.

Easy Operation

The Sievers 860 can operate standalone via the instrument's keypad or with a PC.

Best-in-Class Software

DataPlus enables easy analyzer operation, data management, water quality trending, document creation, and regulatory compliance.

Low Maintenance

Features one-year calibration stability and needs just a few hours of annual preventive maintenance.

Total Organic Carbon*
Linear Range 0.03 ppb to 1.5 ppm TOC
Precision 0.1 ppb or 1%, whichever is greater
Accuracy ± 5% at 500 ppb C of sucrose
Analysis Time 6 minutes
Sample Flow Rate (nominal) Analysis mode:  0.25 mL/min
Fast Flush (between samples):  0.5 mL/min
Sample pH >3
Sampling Method Grab Sampling
Operation PC or standalone operation; one PC can control up to four analyzers
Instrument Specifications*
Power Requirements 100-240 ±10% VAC, 60 W, 50/60 Hz
Fuses T1.6 A, 250 VAC (SloBlo), size 5 x 20 mm (appliance inlet)
Ambient Temperature 10 °C to 40 °C
Normal Operating Environment Intended for indoor use only
Maximum Relative Humidity Up to 95%, non-condensing
Maximum Altitude (known to meet performance specifications) 3,000 m (9,843 ft)
Sparge Gas (if used) Inlet pressure: 30 psig to 50 psig
Sparge Gas Flow Rate 120 mL/min to 150 mL/min
Calibration Stability Typically stable for 12 months
Outputs Serial RS-232 output (2); USB port (1); parallel printer port (1)
Installation/Overvoltage Category II
Safety Certifications CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2
No. 61010-1
Pollution Degree 2
Display Color LCD
Dimensions H: 43.0 cm (16.9 in); W: 21.5 cm (8.5 in); D: 40.5 cm (15.9 in)
Weight 10.5 kg (23.0 lb)

 

* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.

 
 
 