mobile-process
 
 

Process Analysis Systems & Total Solutions for Continuous Water & Wastewater Analysis

Online Hardness & Silica

5500sc Silica Analyzer, 1 channel, reagents included

REQUEST QUOTATION

Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the Series 5000. The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps. Predictive diagnostic tools, including Hach's proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime. No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently. Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.

PROGNOSYS is now available to provide confidence in your instrument readings. PROGNOSYS will monitor and display the reliability of instrument measurement values and identify when upcoming maintenance tasks are due in an easy-to-read color display.

 

90 Days of Continuous Run Time

Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the Series 5000.

Save Time on Maintenance

The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps.

Avoid Downtime

Predictive diagnostic tools, including Hach's proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime.

Clean, Fast and Easy Reagent Change

No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently.

Verify Easily with Hach Lab Products So You Don't Waste Time Second-guessing

Grab Sample In and Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.

Range:

0 to 5000 µg/L as SiO2

Accuracy:

0–500 µg/L: ±1% or ±1 µg/L of reading, whichever is greater; 500–5000 µg/L: ±5%

Repeatability:

± 0.5 µg/L or ± 1 % of reading, whichever is greater

Sample Pressure:

2 to 87 psi (to Preset Pressure Regulator)

Sample Flow Rate:

55 to 300 mL/min

Response Time:

Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature

Lower Limit of Detection (LOD):

0.5 µg/L